Uniformity of dosage units
905
:
meet the requirements.
procedure for content uniformity
1% Sodium lauryl sulfate solution
Dissolve 15 g of sodium lauryl sulfate in 1.5 L of water.
Standard solution
Dissolve an accurately weighed quantity of
USP Bicalutamide RS in a minimum amount of tetrahydrofuran, and dilute quantitatively with
1% Sodium lauryl sulfate solution to obtain a solution having a known concentration of about 0.05 mg per mL.
Test solution
Transfer 1 Tablet to a 100-mL volumetric flask, add about 10 mL of water, and sonicate for approximately 30 minutes. Add about 80 mL of tetrahydrofuran, and sonicate for 30 minutes to complete dissolution of the bicalutamide. Allow to cool to room temperature, and dilute with tetrahydrofuran to volume. Pass this solution through a 0.45-µm suitable filter unit, transfer 10.0 mL of filtrate to a 100-mL volumetric flask, and dilute with 1% Sodium lauryl sulfate solution to volume.
Procedure
Concomitantly determine the UV absorbances of the
Standard solution and the
Test solution with a suitable spectrophotometer at 270 nm, using
1% Sodium lauryl sulfate solution as the blank. Calculate the quantity, in mg, of bicalutamide (C
18H
14F
4N
2O
4S) in the Tablet taken by the formula:
1000C(AU / AS)
in which
C is the concentration, in mg per mL, of
USP Bicalutamide RS in the
Standard solution; and
AU and
AS are the absorbances obtained from the
Test solution and the
Standard solution, respectively.
Limit of 4-amino-2-(trifluoromethyl)benzonitrile
Mobile phase and System suitability solution
Proceed as directed in the Assay.
Standard solution
Dissolve an accurately weighed quantity of
USP Bicalutamide RS in tetrahydrofuran to obtain a solution having a known concentration of about 0.2 mg per mL. Transfer 5.0 mL of this solution to a 50mL volumetric flask, and dilute with
Mobile phase to volume.
Test solution
Transfer an accurately weighed quantity of the powdered Tablet prepared in the Assay preparation, equivalent to about 50 mg of bicalutamide, to a 25-mL volumetric flask. Add about 2 mL of tetrahydrofuran, and allow to stand for 5 minutes. Add about 20 mL of Mobile phase, and sonicate for 10 minutes. Allow to cool to room temperature, and dilute with Mobile phase to volume. Pass the sample through a suitable 0.2-µm filter.
Chromatographic system (see Chromatography
621
)
Proceed as directed in the
Assay, except to use a liquid chromatograph equipped with a 220-nm detector.
Procedure
Separately inject equal volumes (about 10 µL) of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms, and measure the peak responses. Calculate the percentage of 4-amino-2-(trifluoromethyl)benzonitrile in the portion of Tablets taken by the formula:
100(1/1.4)(CS / CU)(rU / rS)
in which 1.4 is the relative response factor for 4-amino-2-(trifluoromethyl)benzonitrile,
CS is the concentration, in mg per mL, of
USP Bicalutamide RS in the
Standard solution; CU is the concentration, in mg per mL, of bicalutamide in the
Test solution based on the label claim;
rU is the peak area for 4-amino-2-(trifluoromethyl)benzonitrile obtained from the
Test solution; and
rS is the peak area for bicalutamide obtained from the
Standard solution: not more than 0.1% of 4-amino-2-(trifluoromethyl)benzonitrile is found.
[noteThe relative retention time for 4-amino-2-(trifluoromethyl)benzonitrile is about 0.4.
]
Assay
Mobile phase
Prepare a mixture of water, tetrahydrofuran, and acetonitrile (65:20:15).
Standard preparation
Dissolve an accurately weighed quantity of
USP Bicalutamide RS in tetrahydrofuran to obtain a solution having a known concentration of about 0.8 mg per mL. Transfer 5.0 mL of this solution to a 100-mL volumetric flask, and dilute with
Mobile phase to volume.
Assay preparation
Grind not fewer than 20 Tablets to a fine powder. Transfer an accurately weighed quantity of powdered Tablets, equivalent to about 50 mg of bicalutamide, to a 100-mL volumetric flask. Add about 50 mL of tetrahydrofuran, and sonicate for at least 10 minutes to complete dissolution. Allow to cool to room temperature, and dilute with tetrahydrofuran to volume. Pass this solution through a suitable 0.45-µm filter. Transfer 4.0 mL of the filtrate to a 50-mL volumetric flask, and dilute with Mobile phase to volume.
Chromatographic system (see Chromatography
621
)
The liquid chromatograph is equipped with a 270-nm detector, and a 5-mm × 12.5-cm column that contains 3-µm packing L1. The flow rate is about 1.5 mL per minute. The column temperature is maintained at 50

. Chromatograph the
System suitability solution, and record the peak areas as directed for
Procedure: the relative retention time for the bicalutamide related compound B peak is about 1.1; the resolution,
R, between bicalutamide and bicalutamide related compound B is greater than 1.9; the tailing factor of the bicalutamide peak is less than 1.3; and the relative standard deviation for replicate injections calculated for the bicalutamide peak is not more than 2.0%.
Procedure
Separately inject equal volumes (about 10 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the areas for the bicalutamide peaks. Calculate the quantity, in percentage of the label claim, of bicalutamide (C
18H
14F
4N
2O
4S) in the portion of Tablets taken by the formula:
100(CS / CU)(rU / rS)
in which
CS is the concentration, in mg per mL, of
USP Bicalutamide RS in the
Standard preparation; CU is the concentration, in mg per mL, of bicalutamide in the
Assay preparation based on the label claim; and
rU and
rS are the peak areas obtained from the
Assay preparation and the
Standard preparation, respectively.